OR
____________________________________________________________
| |
| |
| |
| |
| |
| |
|___________________ ___________________|
| |
| |
| |
| |
| |
| |
| |
| |
|____________________|
Fold the three tabs and wire the edges together. Connect a wire from
the ground of the P15 amplifier to the cage. If you touch the cage
while you look into the microscope, your body will shield the open
side. If you want to move around while recording, lean one of the 20
inch squares you cut out against the front of the cage.
Oscilloscope
The Tek 2201 oscilloscope is listed above as an example of the sort of
scope you want. But don't buy an oscilloscope new unless you're
rolling in money. Electrophysiology is an undemanding application;
you can get something adequate from an electronics surplus store. The
one feature that's worth paying extra for is storage. If you don't
want to buy surplus, call three or four companies and ask about
reconditioned scopes. These are as good as new and come with
warranties, but cost a lot less.
Micromanipulator
The WPI M3301 is more micromanipulator than you need. All you really
need is Z-axis control. If you're strapped for cash, check out some
more catalogs (try Narashige or Newport Optical); you can probably
come up with something for <$500.
Dissecting Scope and Shielding
There's nothing special about the Chinese SM1 microscope except for
its incredible cheapness (in every sense of the word) and, less
obviously, the external power supply for the lamp. You cannot have a
power cable running from outside into the Faraday cage, because the
hum from the power lines will swamp your signal. Thus dissecting
scopes with a built-in power supply will need to be modified. Even
with an external power supply, the light causes problems. First, the
wire from the supply to the scope is an antenna, picking up noise
outside the cage and bringing it in. Solve this by grounding one of
the two wires leading to the scope. Before you do, though, make
absolutely certain you won't cause a short circuit. Unplug the power
supply, then flip on the power switch. Now use an ohmmeter to check
that neither tab on the plug is connected to either of the lines
leading to the bulb. The second problem is that the power supply
itself generates noise. You solve this by turning the light off while
recording. Alternatively, if you want to go deluxe, buy a DC power
supply (about $200), and then you'll be able to look at your worms
while you record.
Leon Avery, (leon@eatworms.swmed.edu)